GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 L
IGHT
S
PEED
* DICOM C
ONFORMANCE
S
TATEMENT
109 of 157
7.6.1 AE Title/Presentation address Mapping
PPS allows for the configuration of the following parameters that pertain to the remote AE.
Remote MPPS AE IP Address:
IP Address used to contact the remote MPPS
AE
Remote AE (HIS/RIS) IP Port:
IP Port used to contact the remote MPPS AE
Remote MPPS AE Title:
Application Entity Title used to contact the
remote MPPS AE
These parameters define where the MPPS requests will be directed. Configuration of
these parameters is performed by GEHC Field Service engineers, using the installation
facilities.
7.6.2 Configurable Parameters
The following are configurable for the PPS Server AE:
Local (PPS Server AE) AE Title (The default is host name of the scanner appended
with string “_PPS”. If the length exceeds 16 characters the AE Title is truncated to a
length of 16.)
Local IP Address
Local IP net-mask
Local IP gateway
The following parameters are configurable, but need change only in case of a system
software upgrade:
Implementation UID
PDU size
Association time-out period
Session time-out period
7.6.3 Configurable DICOM Data
The following DICOM sequences are currently defaulted to be automatically sent in the
PPS N-CREATE and PPS N-SET however PPS can be configured to not send them:
(0008, 1032) Procedure Code Sequence – sent in the PPS N-CREATE. If not sent, the
PPS N-CREATE will contain an empty Procedure Code Sequence.
(0040, 0260) Performed Protocol Code Sequence sent in the PPS N_SET. If not sent,
the PPS N-SET will not contain a Performed Protocol Code Sequence.