IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
39 of 157
In addition to the C-MOVE response status values defined in DICOM part 4 the following status
values will be returned:
• C000
Indicates that an error occurred while retrieving records from the local
database.
• C001
Indicates all other processing error.
• C011
If the Destination AE returns a “Storage Full” condition this status will be
returned. This status will only be sent if the Destination AE returns a
status of A711 and is only applicable if the Destination AE is another
GE Medical Systems product.
2.3.1.3.4.3 Presentation Context Acceptance Criteria
No criterion.
2.3.1.3.4.4 Transfer Syntax Selection Policy
All protocols detailed above in the tables are supported; no priority selection is given.
2.4 Communication Profiles
2.4.1 Supported Communication Stacks (parts 8,9)
DICOM Upper Layer (Part 8) is supported using TCP/IP.
2.4.2 TCP/IP Stack
The TCP/IP stack is inherited from a UNIX Operating System.
2.4.2.1 Physical Media Support
Ethernet v2.0, IEEE 802.3. Auto senses 10/100 Base T Ethernet connection.
2.4.3 Point-to-Point Stack
A 50-pin ACR-NEMA connection is not applicable to this product.
2.5 Extensions / Specializations / Privatizations
2.5.1 Specialized Information Object Definition
Following is a list of additional private attributes defined for a standard CT IOD.
(0008,0008) Value 3: CT Image IOD specific specializations
AXIAL
LOCALIZER
SEGMENTED
REFORMATTED