VCT Vision ELITE Series Especificaciones Pagina 82

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 157
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 81
GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
82 of 157
6.3.1.1.2 Number of Associations
The Worklist Server AE (SCU) will initiate only one DICOM association at a time to perform a
modality worklist query of a single remote AE.
6.3.1.1.3 Asynchronous Nature
Asynchronous mode is not supported. All operations are performed synchronously.
6.3.1.1.4 Implementation Identifying Information
The Implementation UID for this DICOM Implementation is:
ModalityWorklist for LightSpeed implementation UID 1.2.840.113619.6.50
6.3.1.2 Association Initiation Policy
The Worklist Server AE initiates a new association due to an update operation being initiated
from the LightSpeed user interface.
6.3.1.2.1 Real-World Activity: Worklist Query
6.3.1.2.1.1 Associated Real-World Activity
The operator of the system initiates a query for a modality worklist by either opening the
Schedule screen or by opening the Schedule screen and pressing the Update button. The
choice of which of these two behaviors occurs is user configurable. The Worklist Server will
then initiate an association with the remote AE in order to query for the worklist
A user can configure a number of parameters that directly control the worklist query request.
The user can request worklist items that are intended for the scanner the user is working at, all
items that apply to the modality of the scanner the user is working at or all worklist items
available. These selections and their affects on worklist query parameters are given below:
This Scanner:
Modality, (0008,0060) - set to CT
Scheduled Station AE Title, (0040,0001) - set to local AE title
This Modality:
Modality, (0008,0060) - set to CT
Scheduled Station AE Title, (0040,0001) - zero-length (universal
matching)
All Scanners
Modality, (0008,0060) - zero-length (universal matching)
Scheduled Station AE Title, (0040,0001) - zero-length (universal
matching)
The scheduled dates of procedures of interest can be specified for query by selecting a specific
date range. The date ranges available are Today, Days Before Today, Days After Today and All
Days. These selections and their affects on worklist query parameters are given below:
Today: Scheduled Procedure Step Start Date (0040,0002) - set to
YYYYMMDD, where this date is the current date.
Days Before Today Scheduled Procedure Step Start Date (0040,0002) - set to
Vista de pagina 81