GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
114 of 157
Usage in MPPS Instance Attribute Name Tag
Acquisition
with MWL data
Acquisition
without MWL data
Post-Processing
if selected and
present otherwise
from user input
on the scanner
scanner image header
>Retrieve AE Title (0008,0054) Local system
host-name
Local system host-
name
host-name of the
system
>Referenced Image
Sequence
(0008,1140) One item for
each image
created within the
series
One item for each
image created within
the series
One item for each
image generated by
post-processing
>Referenced Standalone
SOP Instance Sequence
(0040,0220) Not sent as part
of image header
Not sent as part of
image header
Not sent as part of
image header
>All other attributes from
Performed Series
Sequence (which Table
F.7.2.1 of DICOM standard
PS3.4 does not explicitly
list)
Not sent as part
of image header
Not sent as part of
image header
Not sent as part of
image header
All other attributes from
Radiation Dose Module and
Billing and Material Code
Module (which Table
F.7.2.1 of DICOM standard
PS3.4 does not explicitly
list)
Not sent as part
of image header
Not sent as part of
image header
Not sent as part of
image header
Note 1:
• When PPS start ( N-CREATE) message is sent, this element will have the
value “IN PROGRESS”
• When PPS end ( N-SET) message is sent, this element will have either
“COMPLETED” or “DISCONTINUED” based on user selection
7.11 USE OF SPECIFIC DICOM DATA
7.11.1 Patient Level
Attribute Name Tag Usage in CT DICOM Image Header
Patient Name (0010,0010) Copied from SPS, if selected
Patient ID (0010,0020) Copied from SPS, if selected
Patient Birthdate (0010,0030) Copied from SPS, if selected
Patient Sex (0010,0040) Copied from SPS, if selected
Referenced Patient Sequence
(0008,1120) Copied from SPS, if selected
Ref. SOP class uid (0008,1150)
Ref. SOP Instance uid (0008,1155)