GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 L
IGHT
S
PEED
* DICOM C
ONFORMANCE
S
TATEMENT
131 of 157
Current Requested
Procedure Evidence
Sequence
(0040,A375) 1C Not Sent
>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3
Pertinent Other Evidence
Sequence
(0040,A385) 1C Not sent
>Include ' Hierarchical SOP Instance Reference Macro' Table C.17-3
Referenced Instance
Sequence
(0008,114A) 1C Not sent
>Include ‘SOP Instance Reference Macro' Table 10-11
>Purpose of Reference
Code Sequence
(0040,A170) 1 Not sent
>>Include 'Code Sequence Macro' Table 8.8-1
9.4 SR Document Content Module
The tables below capture notes concerning the values stored.
9.4.1 TID 10011 – CT Radiation Dose
NL
Rel with
Parent
VT Concept Name
VM
Req
Type
Notes
1
CONT
AINE
R
EV (113701,
DCM, “X-ray
Radiation Dose
Report”)
1 M Sent
2 >
HAS
CONCEPT
MOD
CODE
EV (121058,
DCM,
”Procedure
reported”)
1 M
Code sequence (P5-08000,
SRT, Computed Tomography
X-ray) sent
3 >
INCLU
DE
DTID (1002)
Observer
Context
1-n
M
Observer context sent with
Device context values equal to
Implementation UID. Single
item sent.
4 >
HAS OBS
CONTEXT
DATE
TIME
EV (113809,
DCM, “Start of
X-ray
Irradiation”)
1 M Sent
5 >
HAS OBS
CONTEXT
DATE
TIME
EV (113810,
DCM, “End of
X-ray
Irradiation”)
1 M Sent
6 >
HAS OBS
CONTEXT
CODE
EV (113705,
Accumulation”)
1 M
Code sequence (113014,
DCM, Study) sent
7 >>
HAS UIDR DCID (10001)
1 M Study Instance UID sent