29 of 157
Each C-FIND SCU operation supports an “Association Timer”, “Operation Inactivity Timer” and
“Session Timer” with time out values of 60 seconds, 300 seconds and 60 minutes respectively.
If a “Cancel” or “Refused” status is returned from the Remote AE the association is closed and
the operation terminated.
The DICOM Server AE will parse each matching C-FIND-RSP reply and ignore the entries it
fails to parse. Tables 2.2.1.2.2.2.1-1 - 2.2.1.2.2.2.1-3 shows the various fields that are
requested at the Study, Series, and Image levels of the C-FIND request.
Query results are filtered based on the Modality field. Only CT/MR Screen Save images are
supported.
Table 2.2.1.2.2.2.1-1: Requested Study Level Keys
Description Type Tag Value
Study date R 0008,0020 Zero length
Study time R 0008,0030 Zero length
Patient’s name R 0010,0010 Zero length
Patient ID R 0010,0020 Zero length
Study id R 0020,0010 Zero length
Study Instance UID U 0020,000D Zero length
Study description O 0008,1030 Zero length
Private Creator Identification P 0009,00xx GEMS_IDEN_01
Suite Id P 0009,xx02 Zero Length
Table 2.2.1.2.2.2.1-2: Requested Series Level Keys
Description Type Tag Value
Modality R 0008,0060 Zero length
Series number R 0020,0011 Zero length
Series Instance UID U 0020,000E Series UID
Series description O 0008,103E Zero length
Manufacturer O 0008,0070 Zero length
Images in series O 0020,1002 Zero length
Table 2.2.1.2.2.2.1-3: Requested Image Level Keys
Description Type Tag Value
Image number R 0020,0013 Zero length
Image Instance UID U 0008,0018 Image UID
Image type O 0008,0008 Zero length
Rows O 0028,0010 Zero length
Columns O 0028,0011 Zero length
Image position O 0020,0032 Zero length
Image orientation O 0020,0037 Zero length
Slice thickness O 0018,0050 Zero length
Vista de pagina 28