GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
110 of 157
The configuration is performed by GEHC Field Service Engineers using software
environment variables.
7.7 SUPPORT OF EXTENDED CHARACTER SETS
PPS supports ISO_IR 100 character set.
7.8 N-CREATE & N-SET REQUEST MESSAGE
PPS Feature for LightSpeed supports all named attributes listed in Table F.7.2.1 in PS3.4
of DICOM standard. That is, attributes that are not explicitly referenced by name in the
table are not supported. (Example is last row in the table reads “All other attributes from
Radiation Dose Module and Billing and Material Code Module”. The attributes referenced
here are not supported).
For the MPPS associated with an acquisition, the following attributes are copied from the
Modality Worklist SCU into the MPPS request Message, if procedure performed
corresponds to the SPS information retrieved through the Modality Worklist.
LightSpeed supports the selection of single or multiple SPS for a scan. The following are
applicable.
Single SPS selection results in single PPS message
Multiple SPS selection results in single PPS message
Multiple SPS selection is allowed only if they all correspond to same patient id
A maximum of 15 SPS’s can be selected
Referenced Study Sequence - a maximum of ten Item’s is supported. This attribute will
be present only if SPS information is available from Modality Worklist SCU.
At the end of acquisition the user might choose to ‘Defer PPS’ and later choose to
‘Complete PPS’ or ‘Discontinue PPS’ from the user interface provided in the system.
In this case, the date and time when user chooses to ‘Complete PPS’ or ‘Discontinue
PPS’ is taken as the Performed Procedure Step End Date and Performed Procedure
Step End Time respectively (Not the actual end date and end time of acquisition)
Mapping of SPS data to MPPS SOP instance is explained in section 7.10
Mapping of specific SPS data to CT DICOM IMAGE HEADER , for PPS is explained
in section 7.11
For the MPPS associated with a post-processing the following restrictions apply on the
attributes listed below.
Referenced Study Sequence – The sequence is not sent in the MPPS message
Scheduled Step Attribute Sequence – a maximum of ONE item is supported. The
attribute will be send only if SPS information is available in the image instance.