GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
72 of 157
• Magnification Type (2010, 0060)
• Min Density (2010, 0120)
• Max Density (2010, 0130)
• Empty Image Density (2010, 0110)
• Border Density (2010, 0100)
• Configuration Information (2010, 0150)
• Smoothing Type (2010, 0080)
• The following network timers are supported and are configurable. When these timers expire
the association is terminated. The default values are:
• Association Time out = 120 seconds
• Session Time out = 1200 seconds (total time for open association)
• N-SET Time out = 300 seconds
• N-ACTION Time out = 300 seconds
• N-CREATE Time out = 300 seconds
• N-DELETE Time out = 300 seconds
• N-GET Time out = 100 seconds
• Time-outs, which are set for all hosts, are configurable in dprint.cfg:
*Association time-out - bi_assoc_tio
*Session timeout - bi_session_tio
*Create timeout - bi_ncreate_tio
Note: All configurations are performed by a GE Field Service Engineer.
4.7 Support of Extended Character Set
The Print SCU does not support any Extended Character Set.