GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
136 of 157
Abdomen SRT T-D4000 Abdomen
Spine SRT T-11503 Lumbar spine
Pelvis SRT T-D6000 Pelvis
Lower Extremity SRT T-D0300 Extremity
9.5 CONFIGURATION
The product is configured by default to not generate the CT X-Ray Radiation Dose SR Record. Creation
of the CT Dose Record can be enabled via reconfig.
The product implements the CT Radiation Dose Report template (TID 10011) per DICOM Supplement
127. The DICOM standard says that the SOP class shall be the “X-Ray Radiation Dose SR” SOP Class
(1.2.840.10008.5.1.4.1.1.88.67). However, since many products may not support this SOP class, the
product supports the option via reconfig to create the CT Radiation Dose Report as an “Enhanced SR
SOP Class (1.2.840.10008.5.1.4.1.1.88.22). reconfig “Dose Report:” set to “Full” will produce the X-
Ray Radiation Dose SR SOP class and “Yes” will produce the Enhanced SR SOP class