VCT Vision ELITE Series Especificaciones Pagina 115

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 157
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 114
GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 L
IGHT
S
PEED
* DICOM C
ONFORMANCE
S
TATEMENT
115 of 157
7.11.2 Study Level
Attribute Name Tag Usage in CT DICOM Image Header
Study Instance UID (0020,000D) Copied from SPS, if selected
Study ID (0020,0010) Scanner generated study ID
Referring Physicians name (0008,0090) Copied from SPS, if selected
Accession Number (0008,0050)
If multiple SPS’s are selected, then accession
number from the first selection ( determined by
the user) is used, however, it is configurable to
be set to zero length. Default is to use the
accession number from the first selection.
Referenced Study Sequence (0008,1110) Copied from SPS, if selected
Ref. SOP class uid (0008,1150)
Ref. SOP Instance uid (0008,1155)
7.11.3 Series Level
Attribute Name Tag Usage in CT DICOM Image Header
Modality (0008,0060) Value “CT” is copied
Protocol Name (0018,1030) Name of the selected protocol is copied
Operator Name (0008,1070)
Copied from SPS if selected and present
otherwise generated from user input on scanner
Referenced Performed
Procedure Step Sequence
(0008,1111)
Will be present only if SPS , obtained form
HIS/RIS is selected for scanning
Ref. SOP class uid (0008,1150) Value of MPPS SOP class UID
Ref. SOP Instance uid (0008,1155) Scanner generated unique UID
Requested Attribute
Sequence
(0040,0275)
Will be present only if SPS obtained from
HIS/RIS is selected for scanning. If multiple
SPS’s are selected then this will contain multiple
items one for each SPS. For all other cases this
element will not be present
Requested Procedure Id (0040,1001) Copied from SPS, if selected
Scheduled Procedure
Step Id
(0040,0009) Copied from SPS, if selected
Scheduled Procedure
Step description
(0040,0007) Copied from SPS, if selected
Scheduled Protocol Code
Sequence
(0040,0008) Copied from SPS, if selected
Performed Procedure Step Id (0040,0253) Created at the scanner. The string
PPS_ID_<examnumber>” is used.
Performed Procedure Step
start date
(0040,0244) Same as exam start date
Performed Procedure Step
start time
(0040,0245) Same as exam start time
Vista de pagina 114